| Parameter | Specification |
| Wavelength range | 195nm – 1000nm |
| Wavelength scale interval | 195–250nm: 0.2nm/div |
| 250–350nm: 0.5nm/div | |
| 350–450nm: 1.0nm/div | |
| 450–600nm: 2.0nm/div | |
| 600–1000nm: 5.0nm/div | |
| Wavelength accuracy | 195–340nm: ±1.0nm |
| 340–900nm: ±4.0nm | |
| 900–1000nm: ±4.0nm | |
| Wavelength repeatability | 195–340nm ≤0.5nm |
| 340–1000nm ≤2.0nm |
| Parameter | Specification |
| Transmittance range | 0.0% – 110.0% T |
| Absorbance range | -0.041 – 2.000 A |
| Concentration range | 0 – 9999 C |
| Parameter | Specification |
| Slit width | 0 – 2 mm (continuously adjustable) |
| Timing print range | 3 – 255 seconds |
| Print frequency | 1 – 99 times |
| Parameter | Specification |
| Control system | Single-chip microcomputer |
| Power supply | AC 220V ±22V, 50Hz ±1Hz |
| Lamp system | Integrated with main unit |
| Feature | UV751GD Spectrophotometer | Conventional UV-Vis Instruments |
| Wavelength range | 195–1000nm (UV–Vis–NIR) | Usually 200–800nm |
| System control | Single-chip microcomputer | Analog or basic digital system |
| Lamp integration | Integrated power supply | External lamp system |
| Stability | High stability system | Moderate stability |
| Slit adjustment | Continuously adjustable (0–2mm) | Fixed or limited adjustment |
| Automation | Auto zero, auto full-scale, regression | Manual adjustment required |
| Data output | Timed printing supported | Limited or no automation |