Position: Home > > Analytical instruments > Element analyzer > Oxygen nitrogen hydrogen analyzer > H - 3000 hydrogen analysis apparatus

H - 3000 hydrogen analysis apparatus

Product ID:FYydq004

Price: US$30,000.00 ~ US$80,000.00 PDF Format
Supply Ability:20SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberH - 3000
    Brand NameYIMA
    Payment TermsT/T, paypal
    H - 3000 hydrogen analysis apparatus
    Brief introduction:
    The instrument can be used for metal, alloy and hydrogen content in the determination of inorganic materials materials. Using pulse melting also - the principle of thermal conductivity detection. Under the carrier gas atmosphere, to a certain quality of samples by pulse furnace heated to a certain temperature, hydrogen release in H2 way, of the sample after purification, the measured gas after thermal conductivity detection unit is converted to electrical signals, by integral method to calculate the hydrogen content.
    Technical indicators:
    1. The detection principle, inert gas protection - pulse heating molten - thermal conductivity method.
    2. The relevant standards: GB/T 223.82 2007
    3. The scope of analysis: hydrogen: low hydrogen: 0-0.005% *
    High hydrogen: - 0.1% *
    Note: * 1 g sample, change the sample quantity can expand the measurement range.
    4. The analysis accuracy: standard gas: 0.025 parts per million (0 ~ 1 PPM)
    Standard sample: 0.2 PPM * * or 2%
    Note: * to no greater than the sample standard deviation or uncertainty shall prevail.
    5. Spirit acuity: 0.001 PPM
    6. Analysis time: general for 3 minutes
    7. Sample weighing: general is 1 g, according to the sample content change according to quantity.
    8. The carrier: high purity nitrogen;
    Dynamic strength: ordinary nitrogen or compressed air.
    9. Instrument structure, module structure, by the host, computer, printer, electronic balance, cooling circulating water module composition and so on. Note: * for options.
    10. Detection system: hydrogen analyzer thermal conductivity detection system with high precision.
    A) detector: the antioxidant NTC thermistor element;
    B) signal processing: a small current control technology, to prevent thermal element under the condition of not TongZai gas oxidation;
    C) temperature control: adopt high precision temperature control system;
    D) reference gas path: the good stability of micro flow control;
    11. Flow control: using high precision electronic flow control technology;
    12. Copper oxide catalytic furnace: analysis of gas before entering the detection system, through the copper oxide catalytic furnace, turn the analysis of impurities in gas absorption, guarantee the accuracy of the analysis.
    13. Melting furnace: 0-1500 - a current, power: 8 kva, the highest temperature higher than 3000 ℃.
    14. Correction: two methods
    (1) rapid correction (2) multiple correction results
    15. The power 220 vac? 10%, 50/60 hz, big current 50 a 10 kw (high power).

     
     
  • Brief introduction:
    The instrument can be used for metal, alloy and hydrogen content in the determination of inorganic materials materials. Using pulse melting also - the principle of thermal conductivity detection. Under the carrier gas atmosphere, to a certain quality of samples by pulse furnace heated to a certain temperature, hydrogen release in H2 way, of the sample after purification, the measured gas after thermal conductivity detection unit is converted to electrical signals, by integral method to calculate the hydrogen content.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
18620722006
instruments@instrumentstrade.com