| Parameter | Specification |
| Model | ZDJ-5B-D |
| Instrument Type | Automatic titrator |
| Measurement Modes | Potentiometric and conductivity |
| Display | 7-inch color touchscreen |
| pH Measurement Range | 0.00–14.00 pH |
| mV Measurement Range | −1999.9 to +1999.9 mV |
| Temperature Measurement Range | −5.0°C to +105.0°C |
| pH Resolution | 0.01 pH |
| mV Resolution | 0.1 mV |
| Temperature Resolution | 0.1°C |
| Electronic Unit Error, pH | ±0.01 pH ±1 digit |
| Electronic Unit Error, mV | ±0.03% of full scale |
| Electronic Unit Error, Temperature | ±0.3°C |
| Burette Capacity Options | 10 mL and 20 mL |
| 10 mL Burette Capacity Tolerance | ±0.025 mL |
| 20 mL Burette Capacity Tolerance | ±0.035 mL |
| Titration Analysis Repeatability | 0.2% |
| Electronic Unit Repeatability Error | ≤0.2 mV |
| Electronic Unit Stability | ±0.3 mV ±1 digit / 3 h |
| Titration Control Sensitivity | ±2 mV |
| Electrode Calibration | Automatic three-point pH calibration |
| Buffer Recognition | GB, DIN, and NIST |
| Method Storage | Up to 100 titration methods |
| Result Storage | Up to 200 titration results |
| Communication | USB and RS232 |
| Computer Software | Compatible titration software available |
| User Management | Password management; up to 8 user accounts |
| Operating Languages | Chinese and English |
| Comparison Item | ZDJ-5B-D Automatic Titrator | Conventional Manual Titration |
| Titrant Delivery | Automated, controlled delivery | Typically controlled manually |
| Endpoint Determination | Electrode-based measurement and method evaluation | Often based on visual indicators or manual readings |
| Titration Curve | Real-time curve display | Usually requires separate recording or analysis |
| Method Management | Store, copy, edit, and create methods | Often relies on written procedures |
| Electrode Calibration | Automatic three-point pH calibration | Depends on separate calibration procedures |
| Data Storage | Up to 200 titration results | Depends on manual or external records |
| Computer Connectivity | USB and RS232 | Depends on additional equipment |
| Multi-Sample Automation | Autosampler support available by configuration | Usually requires repeated manual handling |
| Best Suited For | Standardized workflows and electronic data management | Basic titration tasks and low-throughput applications |