| Parameter |
Specification |
| Model |
Y09-3016 |
| Product Type |
Handheld Laser Dust Particle Counter |
| Light Source |
Semiconductor laser, lifetime >35,000 hours |
| Sampling Flow Rate |
2.83L/min (0.1ft³/min) |
| Particle Size Channels |
0.3, 0.5, 1.0, 3.0, 5.0, 10μm (6 channels) |
| Counting Modes |
Cumulative value, differential value, concentration value |
| Measurement Modes |
Single, repeat, continuous, calculation, remote |
| Sampling Period |
1-10000s |
| Delay Counting |
0-99s |
| Self-cleaning Time |
≤10min |
| Display |
LCD real-time display |
| Display Data |
Particle concentration, time, date, RH/temp, room number, sampling point, battery status |
| Particle Concentration Unit |
Particles/m³ and unit conversion |
| Battery |
Rechargeable lithium-ion battery, 7.4V, 2200mAh |
| Working Time |
Approx. 6 hours |
| Power Supply |
AC100-245V, 50/60Hz adapter to DC7.4V 1A |
| Data Storage |
1000 measurement groups |
| Communication Interface |
RS232, 9600 baud rate |
| Alarm Function |
Cleanroom level alarm and low battery alarm |
| Zero Count |
≤1 particle/5 min |
| Overlapping Error |
≤5% at 2,000,000 particles/ft³ |
| Operating Temperature |
10-40℃ |
| Operating Humidity |
20-90%RH (non-condensing) |
| Atmospheric Pressure |
86-106KPa |
| Optional Temperature Range |
0-50℃ ±1℃ |
| Optional Humidity Range |
0-100%RH ±5% |
| Dimensions |
230×130×45mm |
| Weight |
0.6kg |
| Calibration |
Traceable to NIST standards |
| Standard Accessories |
Carrying case, charger, isokinetic sampling head, zero filter, RS232 cable, software |
| Optional Accessories |