| Item | Specification |
| Product Model | SZC-D |
| Measurement Principle | Soxhlet extraction and gravimetric determination |
| Sample Capacity | 6 samples simultaneously |
| Measurement Range | Crude fat content ≥0.5% |
| Applicable Samples | Grain, food, feed, oilseeds and various oil products |
| Power Supply | 220 V ±10%, 50 Hz |
| Power Consumption | 500 W |
| Temperature Range | Room temperature to 200°C |
| Temperature Adjustment | Continuously adjustable |
| Temperature Control Accuracy | ±1°C |
| Solvent Recovery Rate | ≥80% |
| Extraction Recovery Rate | ≥99% |
| Accuracy | Relative difference ≤3% |
| Parallel Difference | ≤0.3% |
| Heating Method | Fully enclosed aluminum alloy electric heating plate |
| Sealing Method | PTFE sealing |
| Reflux Rate | 120 drops/min |
| Cooling Water Temperature | ≤30°C |
| Dimensions | 590 × 390 × 625 mm |
| Weight | 25 kg |
| Timing Function | Yes |
| End-of-Test Alarm | Yes |
| Automatic Solvent Recovery | Yes |
| Feature | SZC-D Fat Analyzer | Conventional Manual Extraction |
| Extraction Principle | Soxhlet extraction | Soxhlet or manual solvent extraction |
| Sample Capacity | 6 samples simultaneously | Usually lower or more operator-dependent |
| Temperature Control | Automatic digital control | Manual or less automated |
| Temperature Range | Room temperature–200°C | Depends on heating equipment |
| Solvent Recovery | Automatic | Usually requires additional operation |
| Timing | Automatic timing | Manual timing |
| End-of-Test Alarm | Available | Usually unavailable |
| Heating | Fully enclosed aluminum alloy heating plate | Conventional heating system |
| Sealing | PTFE sealing | Depends on setup |
| Operation | Integrated and simplified | More manual procedures |
| Laboratory Efficiency | High | Relatively lower |