| Parameter | Specification |
| Model | SN-T |
| Product Type | Benchtop Xenon Arc Weathering Test Chamber |
| Temperature Range | RT+10℃~70℃ |
| Black Panel Temperature | 63℃±3℃ |
| Temperature Fluctuation | ≤±0.5℃ |
| Temperature Uniformity | ≤±2℃ |
| Temperature Deviation | ≤±1.5℃ |
| Sample Tray Size | 251 × 460 mm |
| Glass Filter | 1 piece |
| Xenon Lamp Type | Air-cooled xenon lamp |
| Xenon Lamp Quantity | 2 pcs (1 operating + 1 spare) |
| Xenon Lamp Power | 2.5KW × 1 |
| Sample Rack to Lamp Distance | 230~280 mm |
| Spray Cycle | 102 min light / 18 min spray or 48 min light / 12 min spray |
| Spray Pressure | 0.12~0.15 MPa |
| Spray Nozzle Diameter | Φ0.8 mm |
| Power Supply | AC380V / 50Hz |
| Irradiance Mode | Value |
| Broadband 300–400 nm | 60 W/m² |
| Narrowband 420 nm | 1.10 W/m² |
| 340 nm Irradiance | 0.51 W/m² |
| Full Spectrum 290–800 nm | 550 W/m² |
| Industry | Testing Applications |
| Automotive Industry | Interior materials, exterior parts, coatings, plastics |
| Coating Industry | Paints, protective coatings, pigments |
| Rubber Industry | Rubber products, seals, tires |
| Plastic Industry | Polymer aging resistance evaluation |
| Textile Industry | Fabric color fastness and durability testing |
| Adhesive Industry | Aging resistance of bonding materials |
| Research Institutions | Material development and reliability studies |
| Comparison | Xenon Arc Weathering Test Chamber | UV Weathering Tester |
| Light Source | Xenon arc lamp | UV fluorescent lamp |
| Spectrum Simulation | Full solar spectrum | Mainly UV spectrum |
| Testing Accuracy | Closer to real outdoor exposure | Focuses on UV degradation |
| Applicable Materials | Wide range of materials | Mainly UV-sensitive materials |
| Applications | Automotive, plastics, coatings, textiles | Coatings, polymers, outdoor materials |
| Environmental Simulation | Light + heat + humidity + spray | Mainly UV + condensation |