| Parameter | Specification |
|---|---|
| Measurement Range | 0.0001–0.1% (m/m) |
| Potential Measurement Range | −1999.9 to +1999.9 mV |
| Electronic Unit Basic Error | 0.1% F.S. ±0.1 mV |
| Input Impedance | ≥1×10¹² Ω |
| Burette Volume | 10 mL |
| Minimum Titration Volume | 0.001 mL |
| Burette Accuracy | ±0.1% F.S. |
| Precision | Complies with GB/T 1792 |
| Compatible Measurements | Acid value, base value, basic nitrogen in dark oils, pH measurement, etc. |
| Ambient Temperature | 0–35°C |
| Power Supply | AC 220±22 V, 50 Hz |
| Feature | MIA-4AJ Mercaptan Sulfur Analyzer | Manual Potentiometric Titration |
|---|---|---|
| Titration Control | Automatic | Manual |
| Endpoint Determination | Automatic + manual correction | Mainly manual |
| E-V Curve | Real-time display | Usually requires separate recording |
| dE/dV Curve | Supported | Limited/manual |
| Data Storage | Large-capacity digital storage | Manual or separate software |
| Curve Processing | Zoom, move and correction | Limited |
| Automatic Cleaning | Yes | No |
| Automatic Replenishment | Yes | No |
| Data Export | Excel/Word | Usually manual |
| Printing | Automatic results and curves | Usually separate process |