| Mode | Function |
|---|---|
| DET | Dynamic Equivalence Point Titration with preset initial volume |
| MET | Monotonic Equivalence Point Titration |
| SET | Preset Endpoint Potential Titration |
| CAL | pH Calibration |
| MEAS | pH Measurement |
| Parameter | Specification |
|---|---|
| mV Measurement Range | −2000.0 to +2000.0 mV |
| pH Measurement Range | 0–20.000 pH |
| mV Resolution | 0.1 mV |
| pH Resolution | 0.001 pH |
| Electronic Unit Error, mV | ±0.03% F.S. ±1 digit |
| Electronic Unit Error, pH | ±0.001 pH ±1 digit |
| Electronic Unit Input Impedance | ≥3 × 10¹² Ω |
| Electronic Unit Input Current | ≤1 × 10⁻¹² A |
| Electronic Unit Stability | ±0.2 mV ±1 digit / 8 h |
| Burette Volume | 10 mL |
| Minimum Delivery Volume | 0.001 mL |
| Burette Capacity Accuracy | ±0.010 mL |
| Burette Filling/Delivery Speed | 55 ±5 s/full burette |
| Titration Repeatability Error | ≤0.2% |
| Reference Titration Condition | 0.1 mol/L HCl titrating 0.1 mol/L NaOH |
| Display | 6-inch large LCD |
| Control | Microcomputer control |
| Interface | Mouse-based human-machine interaction |
| Feature | DDY-2008J Automatic Potentiometric Titrator | Conventional Manual Titration |
|---|---|---|
| Titrant Addition | Automatic | Manual |
| Endpoint Detection | Automatic + manual options | Usually visual/manual |
| Titration Curve | Real-time E-V and DE/DV | Usually unavailable |
| False Endpoint Filtering | Supported | Manual judgment |
| Multi-Endpoint Titration | Supported | Manual |
| Cleaning | Automatic | Manual |
| Replenishment | Automatic | Manual |
| Fixed-Volume Dosing | Automatic | Manual |
| Result Calculation | Automatic | Manual calculation often required |
| Data Recording | Automatic | Manual |
| Data Printing | Results + curves + complete data | Usually manual |
| Titration Modes | DET, MET, SET, CAL, MEAS | Depends on operator |
| Electrode Inputs | Dual high-impedance inputs | Manual measurement |
| Non-Aqueous Titration | Supported | Operator dependent |