| Parameter | Specification |
| Configuration | 8 stations (4 front, 4 rear) |
| Standard Accessories | 8 paddles and 8 vessels |
| Paddle Shaft Swing | ≤0.3 mm |
| Basket Swing | ≤0.8 mm |
| Shaft-to-Vessel Axis Deviation | ≤1 mm |
| Rotation Speed Range | 20–200 rpm |
| Speed Resolution | 1 rpm |
| Water Bath Temperature Range | Room temperature to 45°C |
| Temperature Resolution | 0.1°C |
| Temperature Control Accuracy | ≤±0.3°C |
| Preset Sampling Periods | 9 programmable periods |
| Sampling Period Duration | 1–999 min per period |
| Timing Error | ±0.5 min |
| Stored Pharmacopoeia Parameters | 300 Chinese Pharmacopoeia dissolution test parameter sets |
| Test Head Operation | Automatic electric lifting and tilting |
| Temperature Monitoring | Electronic sensors monitor water bath temperature at multiple points |
| Heating Power | 1,100 W |
| Optional Accessories | Small-vessel method accessories |
| Comparison Item | ZRS-8ST | Conventional Manual-Access Dissolution Tester |
| Test Capacity | 8 stations | Depends on instrument configuration |
| Test Head Operation | Automatic electric lifting and tilting | May require manual handling |
| Parameter Storage | Automatic parameter saving | May require repeated setup |
| Temperature Control | Integrated sensor and software correction | Depends on control system |
| Sampling Schedule | 9 programmable sampling periods | Depends on instrument functionality |
| Pharmacopoeia Parameters | 300 stored Chinese Pharmacopoeia parameter sets | Depends on built-in database |
| Safety Functions | Fault alarms and over-temperature protection | Depends on instrument configuration |