| Parameter |
Specification |
| Model |
UPW-OB100 |
| Instrument Type |
Portable Fully Automatic Infrared Oil Content Analyzer |
| Measurement Principle |
Infrared absorption spectroscopy |
| Baseline Stability |
Real-time automatic zero correction; designed to eliminate baseline drift effects |
| Method Detection Limit |
0.0125 mg/L (1,000 mL sample, 25 mL extract, 40 mm optical path) |
| Method Detection Limit |
0.05 mg/L (500 mL sample, 50 mL extract, 40 mm optical path) |
| Minimum Detection Concentration |
0.0006 mg/L, as specified by the manufacturer |
| Basic Measurement Range |
0–100 mg/L |
| Maximum Measurement Concentration |
Up to 100% oil, with automatic dilution for high concentrations |
| Accuracy Error |
≤2% |
| Repeatability |
RSD ≤1% for 11 measurements of an oil sample at 30–40 mg/L |
| Linear Correlation Coefficient |
R > 0.999 |
| Wavenumber Range |
60,000–85,000 m⁻¹, equivalent to approximately 2940–4167 nm |
| Wavenumber Accuracy |
±50 m⁻¹, as interpreted from the supplied specification |
| Wavenumber Repeatability |
±25 m⁻¹, as interpreted from the supplied specification |
| Absorbance Range |
0.00000–2.00000 AU |
| Corresponding Transmittance Range |
100%–1% T |
| Full-Spectrum Scanning Speed |
30 seconds per scan |
| Non-Dispersive Infrared Measurement Time |
2 seconds per measurement |
| Overall Sample Measurement Time |
≤10 minutes |
| Sample Volume Range |
0–600 mL |
| Sample Volume Measurement |
Automatic, with error ≤2% |
| Measurement Modes |
Full-spectrum scanning, three-wavenumber measurement, and single-point non-dispersive infrared |
| Calibration Methods |
Three-standard correction, standard-curve calibration, and single-point calibration |
| Automated Functions |
Demulsification, extraction, water removal, measurement, data storage, and result printing |