    Suxin sx-bhc series biosafety cabinet is a biosafety isolation device used in biosafety laboratories or other laboratories for the protection of personnel, test samples and the environment, which can meet the risk level of 1, 2, 3, 3 disease body operations.
    SX - BHC series biological safety cabinet is biological safety cabinet, the front opening area to the negative pressure suction in the airflow to protect the staff's safety, the high efficiency filter (HEPA) vertical airflow is used to protect the safety of the sample subjects, vent gas flowing through the high efficiency filter, protect the environment from pollution.
    Performance features:
    1. The box adopts high-quality SUS304 stainless steel or cold-rolled steel plate, which is easy to clean;
    2. Vertical laminar negative pressure type, with 30%/100% filtered discharge of return air;
    3. Negative pressure design to ensure the safety of operators;
    4. Integrated sliding security door positioning, easy to operate;
    5. Equipped with filter blocking alarm, fan overload alarm, and working window opening limit alarm system.
    Technical parameters:
    Model no. : SX-BHC-1300A2  SX-BHC-1300B2
    Clean grade: ISO5 (100) /ISO4 (10)
    Filter level/filter efficiency: HEPA/ULPA@0.3-0.12um greater than or equal to 99.999%-99.9995%
    Average flow velocity of downdraft (m/s) : 0.5-0.5m /s
    Average flow velocity (m/s) of incoming airflow is greater than or equal to 0.5m/s
    Noise: 67dB (A)
    Vibration half-peak: less than 5um
    Power supply: AC single-phase 220V/50HZ
    Biological protection of air flow balance:
    Staff protection 1-8 cfu /ml (repeat 3 times, 5min/ time)
    A. Total number of colonies of impact sampler: less than or equal to 10CFU/ time
    B. Total number of colony of slit sampler: less than 5CFU/ time
    Product protection 1-8 cfu /ml (repeat 3 times, 5min/ time)
    Total number of colonies: less than 5CFU/ time
    Cross-contamination protection 1-8 cfu /ml (repeat 3 times, 5min/ time)
    Total number of colonies: less than or equal to 2CFU/ time
    Maximum power (KVA) : no more than 1500VA (including standby socket)
    Weight (kg) : 200
    Working area size W*D*H (mm) : 1300*570*635
    Device profile size W*D*H (mm) : 1500*855*2350
    Specification and quantity of fluorescent lamp: 40W*2
    Specification and quantity of uv lamp: 30W*1
    Illumination: 650 lx
    Send, exhaust duct diameter: 250 mm/Ø Ø 250 mm
    Direction of wind: push out, push out

     
  • Product description:
    Suxin sx-bhc series biosafety cabinet is a biosafety isolation device used in biosafety laboratories or other laboratories for the protection of personnel, test samples and the environment, which can meet the risk level of 1, 2, 3, 3 disease body operations.
    SX - BHC series biological safety cabinet is biological safety cabinet, the front opening area to the negative pressure suction in the airflow to protect the staff's safety, the high efficiency filter (HEPA) vertical airflow is used to protect the safety of the sample subjects, vent gas flowing through the high efficiency filter, protect the environment from pollution.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
18620722006
instruments@instrumentstrade.com