PSD-11 electrolyte analyzer

Product ID:SHLCdjz 013

Price: Please inquire PDF Format
Supply Ability:20SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Brand NameNJPSDDZYQ
    Payment TermsT/T, paypal
    Model NumberPSD-11
    1. Basic parameters:
    1.1 measurement principle: ion selective electrode + gas sensor
    1.2 measurement method: direct measurement method
    1.3 sampling method: manual (psd-11/15a /16A), automatic (psd-15b /16B)
    1.4 calibration mode: automatic calibration
    1.5 measurement items: K+, Na+, Cl-, Ca++, Ph, TCO2 (except psd-15a /15B)
    1.6 measurement range and display resolution:
    analysis item measurement range Display resolution
    K+ 0.50 mmol/L 15.0 mmol/L 0.01 mmol/L
    Na+ 30.0 mmol/L 200.0 mmol/L 0.1 mmol/L
    Cl- 30.0 mmol/L 200.0 mmol/L 0.1 mmol/L
    Ca++ 0.10 mmol/L 5.00 mmol/L 0.01 mmol/L
    TCO2 5.0mmol/L60.0mmol/L 0.1mmol/L
    pH 6.009.00 0.01
    1.7 type of sample: serum, plasma, whole blood, diluted urine (except Ca++, TCO2, pH)
    1.8 analysis speed: 40 samples /1h (11, 16A, 16B), 90 samples /1h (15A, 15B)
    1.9 the maximum sample size: less than or equal to 200 for a person (11, 16A, 16B) and 100 for a person (15A, 15B)
    2. Performance indicators:
    terms of the name accuracyV precisionCV linearΔ stabilityΔD Cross contamination rateΔξ
    ask K+ ≤0.3mmol/L ≤1.0% ≤3.0% ≤2.0% ≤1.5%
    Na+ ≤3.0mmol/L ≤1.0% ≤3.0% ≤2.0% ≤1.5%
    Cl- ≤3.0mmol/L ≤1.0% ≤3.0% ≤2.0% ≤1.5%
    Ca++ ≤0.3mmol/L ≤1.5% ≤5.0% ≤2.0% ≤1.5%
    TCO2 ≤3.0mmol/L ≤3.0% ≤5.0% ≤5.0% ≤2.5%
    pH —— ≤1.0% —— —— ——
    3. Environmental parameters:
    Normal working conditions of the instrument:
    A) power voltage: AC220V;
    50 hz.
    B) the environmental temperature 5 to 40 ;
    C) relative humidity: less than or equal to 80%;
    D) atmospheric pressure: 760hPa ~ 1060hPa;
    E) keep away from strong electromagnetic interference sources;
    F) avoid direct exposure to strong light;
    G) good grounding environment.
    4. Electrical parameters:
    A) power supply: AC220V;
    50 hz.
    B) input power: 65VA;
    C) fuse: T1AL250V, Φ 5 x 20.
    5. Functional parameters:
    A) operation mode: all-chinese menu operation;
    B) program functions: built-in real clock, parameter setting, maintenance, quality control, fault diagnosis and other functions;
    C) display: 128 x 64 (ps-11), 320 x 240 graphic liquid crystal display;
    D) printer: built-in thermal printer;
    E) data storage: samples and quality control test results can be stored;
    F) communication interface: rs-232c serial interface.
    6. Volume and weight:
    A) volume (H * W * D) : 30cm * 25cm * 20cm (psd-11), 40cm * 36cm * 40cm (psd-15a /15B),
    40 cm by 50 cm x 40 cm (16 PSD - 16 a/b)
    B) weight: about 5kg (psd-11), 8kg (psd-15a /15B), 12kg (psd-16a /16B)
    7. Technical features:
    Uppercase two-channel measurement system, simultaneous measurement of ISE and TCO2;
    Working high quality ISE+ gas sensor measurement system, direct sample measurement, completely free of maintenance, with a life of over 1 year.
    The results of K+, Na+, Cl-, Ca++ +, pH and TCO2 in the samples can be determined within 60s after one infusion.
    The latest microprocessor system is adopted, and all operation processes of the instrument, including calibration, sample testing, and sensor status monitoring, are controlled by the program.
    The instrument USES large-screen liquid crystal display and a new operating interface. It helps the operator to use the instrument conveniently and quickly through multiple shortcut function keys. Various information of the instrument is displayed on the screen at the same time.
    The superior electrode state tracking and tuning system ensures reliability of test results at all times.
    Working full electromagnetic valve sealed liquid circuit control system, the sample channel is automatically rinsed throughout the whole process, which is not suitable for plugging the hole. Meanwhile, the cross contamination in the sample measurement is reduced to extremely low level.
    Superior optical sample detector is used to completely eliminate incorrect measurement results.
    Superior quality control procedures are set up for calculating and storing the quality control data of all 3 quality control groups within the last 31 days, and any two of them can be used to correct the instrument.
    Working complete data management system can store samples and quality control test results, and it has the function of power-off protection and automatic refresh when full.
    The data interface of the instrument can be compatible with any type of LIS/HIS system.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
18620722006
instruments@instrumentstrade.com