Position: Home > > Environmental monitoring and analysis > Air sampling equipment > Smoke / flugas sampler >  Model 3072 Intelligent Dual-Channel Flue Gas Sampler

 Model 3072 Intelligent Dual-Channel Flue Gas Sampler

Product ID:HQCYyq046

Price: Please inquire PDF Format
Supply Ability:50 SETS/MON
Port:SHENZHEN
  • Product Introduction
  • Other information
  • Model Number3072
    Brand NameQDLAOYING
    Payment TermsTT,PAYPAL,Alibaba.com Pay
     Model 3072 Intelligent Dual-Channel Flue Gas Sampler
    Introduction:
    This instrument employs the solution absorption method to collect harmful gases such as SO2 and NOX from pollution sources. It is widely used in environmental protection, testing companies, industrial enterprises (power plants, steel mills, cement plants, sugar mills, paper mills, smelters, ceramic factories, boilers and kilns, as well as aluminum, magnesium, zinc, titanium, silicon, pharmaceutical, fertilizer, chemical, rubber, and material industries), public health, labor, safety supervision, military, scientific research, and education sectors.
    Execution Standard:
    GB/T 16157-1996 Determination of Particulate Matter and Sampling Methods for Gaseous Pollutants in Exhaust from Fixed Pollution Sources
    HJ/T 47-1999 Technical Specifications for Flue Gas Samplers
    HJ/T 397-2007 Technical Specification for Exhaust Emission Monitoring of Fixed Sources
    JJG 1169-2019       Flue Gas Sampling Apparatus
    Features:
    1. Dual-channel flexible sampling with independent control for each channel.
    2. Synchronous detection of dynamic pressure, static pressure, total pressure, flue gas temperature, flue gas velocity, flue gas flow rate, and standardized dry flow rate
    3. Excellent flow stability, strong load capacity, ensuring reliable long-term operation
    4. Equipped with functions for storing, querying, printing, and exporting test data
    5. Built-in lithium battery, enabling prolonged use without external power
    6. Optimize the operation interface to better suit users' habits
    Technical Parameters:
    Sampling flow rate: (0.2~1.5) L/min
    Sampling time: arbitrarily set within 99h59min
    Sampling frequency: (1 to 99) times
    Temperature before flow meter: (-55~125)℃
    Pressure before the flow meter: (-25~0) kPa
    Atmospheric pressure: (50~130) kPa
    Smoke dynamic pressure: (0~2000) Pa
    Flue gas static pressure: (-30~+30) kPa
    Flue gas temperature: (0~500)°C
    Flue gas velocity: (1~45) m/s
    Instrument noise: ≤65dB(A)
    External dimensions (length × width × height): 300mm × 180mm × 320mm
    Host weight: approximately 6kg
    Power consumption: <50W
    Power supply: Built-in lithium battery or AC power (with integrated charging management)

     
     

     
  • Introduction:
    This instrument employs the solution absorption method to collect harmful gases such as SO2 and NOX from pollution sources. It is widely used in environmental protection, testing companies, industrial enterprises (power plants, steel mills, cement plants, sugar mills, paper mills, smelters, ceramic factories, boilers and kilns, as well as aluminum, magnesium, zinc, titanium, silicon, pharmaceutical, fertilizer, chemical, rubber, and material industries), public health, labor, safety supervision, military, scientific research, and education sectors.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com