| Parameter | Specification |
| Model | MDS-2014DS |
| Dissolution Vessel Number | 14 vessels |
| Paddle/Basket Number | 14 |
| Independent Operation Mode | 7 × 2 parallel operation |
| Compatible Methods | Basket method, Paddle method, Small vessel method, Paddle over disk method, Rotating cylinder method |
| Heating Range | Room temperature ~45.0℃ |
| Temperature Resolution | 0.1℃ |
| Temperature Accuracy | ±0.2℃ |
| Stirring Speed Range | 10.0–300.0 RPM |
| Speed Resolution | 0.1 RPM |
| Speed Accuracy | ±0.5 RPM |
| Light Protection Function | White light / Red light / Three-level illumination |
| Storage Methods | More than 100 experimental methods |
| Data Transfer | USB import/export |
| User Management | Four-level permission control |
| Printing Function | Operation logs and audit records printing |
| Parameter | Specification |
| Shaft Swing | ≤0.5 mm |
| Basket Swing | ≤1.0 mm |
| Shaft and Vessel Axis Deviation | ≤1.5 mm |
| Instrument Level Accuracy | ≤0.2° |
| Shaft Verticality | 90° ±0.5° |
| Vessel Verticality | 90° ±1° |
| Parameter | Specification |
| Sampling Channels | 14 channels |
| Sampling Volume | 0.1–10 mL (single sample); 0.2–20 mL (dual sample) |
| Sampling Frequency | 1–32 times (single); 1–16 times (dual) |
| Sampling Accuracy | ≤0.8% (10 mL sampling) |
| Automatic Dilution Accuracy | ±1% |
| Minimum Sampling Interval | First sampling: 1 min 20 s; subsequent sampling: 2 min |
| Sampling Functions | Rinsing, filling, discarding, sampling, replacement, dilution, drainage, cleaning |
| Communication Interface | RS485 / 100M RJ45 |
| Sampling Pump | High precision syringe pump |
| Feature | Traditional Dissolution Tester | MDS-2014DS Automatic System |
| Vessel Capacity | Usually 6 or 8 vessels | 14 vessels |
| Sampling Operation | Manual sampling | Fully automatic sampling |
| Drug Addition | Manual operation | Automatic dosing module |
| Temperature Monitoring | Single monitoring | Independent 14-channel monitoring |
| Data Management | Limited recording | Audit trail and user management |
| Cleaning Process | Manual cleaning | Automatic cleaning system |
| Method Transfer | Manual setup | USB import/export |
| GMP Compliance | Basic | Designed for GMP environments |