| Type of analysis station | Micropore analysis unit | Mesoporous analysis unit |
| Measurement range | Specific surface area: ≥ 0.0005 m2/g (RSD ≤ 0.5%, standard sample) Pore size: 0.35-200 nm (maximum pore size repetition deviation ≤ 0. 02 nm, standard sample) Pore volume: ≥ 0. 0001 m3/g |
Specific surface area: ≥ 0.0005 m2/g (RSD ≤ 0.5%, standard sample) Pore size: 0.35-200 nm (the most probable pore size repeat deviation ≤ 0. 2 nm, standard sample) Pore volume: ≥ 0. 0001 m3/g |
| Pressure transducer | 1000 Torr + 10 Torr + 1/0. 1 Torr | 1000 Torr |
| Vacuum system | Imported mechanical vacuum pump + imported molecular pump (ultimate vacuum 10-8Pa) | Imported mechanical vacuum pump (ultimate vacuum 10-2Pa) |
| Range of relative pressure | 10-8 to 0.998 | 10-4 to 0.998 |
| Degassing system | In-situ degassing, each unit is equipped with a heating furnace, and the temperature range is normal temperature-400 ℃ (higher temperatures can be customized) | |
| Piping system | Air bath circulation heating temperature control, the maximum constant temperature can be 45 ℃ | |
| Air intake system | Each working unit is equipped with an independent adsorption gas interface (nitrogen or other gas); Can be equipped with multi-channel air intake device to realize one-button switching of multiple gases | |
| Adsorbed gas | Conventional gases: N2, Ar, CO2, O2, Kr, etc.; Combustible gas: CH4, H2, C2H6, etc. | |
| Basic information | |
| Operating temperature | 15℃-40℃ |
| Operating humidity | 20% RH-80% RH |
| Dimensions and Quality | Single working unit with cold trap : L 70 cm × W 70 cm × H 105 cm, 110 kg; Two working units are cascaded with cold trap : L 105 cm × W 70 cm × H 105 cm, 170 kg; Three working units cascaded with cold traps : L 142 cm × W 70 cm × H 105 cm, 220 kg; |
| power supply | Voltage : AC 220 ± 20 V, 50/60 Hz; Maximum power: 3500 W |