GIT-MINI02 Glove Integrity Tester

Product ID:SHGCglq037

Price: Please inquire PDF Format
Supply Ability:50 SETS/MON
Port:SHENZHEN
  • Product Introduction
  • Other information
  • Model NumberGIT-MINI02
    Brand NameZJTAILIN
    Payment TermsTT,PAYPAL,Alibaba.com Pay
    GIT-MINI02 Glove Integrity Tester
    Introduction:
    The GIT-MINI02 Glove Integrity Tester is a specialized instrument designed to test the integrity of isolators/RABS system sleeves/gloves or integrated gloves. The glove integrity tester features an intelligent integrated control and display, and comes with a test port that matches the isolator's operating port. The tester can operate in positive pressure mode, eliminating the need for manual operation by the operator. The display screen automatically displays the test results throughout the entire process, allowing for the determination of the seal integrity of the tested gloves.
    Features:
    1. The detection process is performed online, without the need to disassemble the cuff/glove.
    2. With an ingenious structure and a portable design, it is convenient to use.
    3. Battery-powered design, no external power supply required.
    4. The tester integrates an internal air supply device, eliminating the need for an external air source.
    5. Controlled by a single-chip microcomputer, 4.3-inch touch screen display.
    6. It has a good human-computer interaction interface, which is convenient for observation and operation.
    7. Used for integrity testing of sleeves/gloves before or after use in isolators.
    8. Developed in accordance with ISO14644-7 standard.
    9. Fully automated monitoring of pressure within the inflatable sealing ring and gloves during the testing process;
    10. It features a variety of formula options to meet diverse glove testing needs;
    11. Glove number identification: RFID chip;
    12. Support three-level operation permissions;
    Technical Specifications:
    Test port specification: Ф255mm (other specifications can be customized)
    Input pressure: > 0.2 MPa
    Inflation pressure: 500Pa~2500Pa
    Test duration: 3 minutes to 8 minutes
    Pressure display resolution: 0.1 Pa
    Detection accuracy (minimum aperture): 300μm;
    Instrument weight: approximately 2.5kg
    Overall dimensions: 260mm×280mm×125mm
    Power supply battery: lithium battery
    Charging time: 4 hours;
    Power: ≤20W
    Continuous working time: ≥4 hours;
    Data processing methods: Online micro-typing, USB export;
    Storage data: 10,000 sets;

     
     
     
     
  • GIT-MINI02 Glove Integrity Tester
    Introduction:
    The GIT-MINI02 Glove Integrity Tester is a specialized instrument designed to test the integrity of isolators/RABS system sleeves/gloves or integrated gloves. The glove integrity tester features an intelligent integrated control and display, and comes with a test port that matches the isolator's operating port. The tester can operate in positive pressure mode, eliminating the need for manual operation by the operator. The display screen automatically displays the test results throughout the entire process, allowing for the determination of the seal integrity of the tested gloves.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com