| Parameter | Specification |
| Model | ET-3015E |
| Measurement Principle | Non-Dispersive Infrared (NDIR) Gas Analysis |
| Target Gas | Carbon Monoxide (CO) in Air |
| Sampling Method | Pump-sampling |
| CO Measurement Range | 0–50 ppm; optional 0–200 ppm or 0–1000 ppm ranges |
| Temperature Range | -20°C to +60°C |
| Humidity Range | 10–95% RH |
| Resolution | 0.1 × 10⁻⁶ |
| Linearity Error | ≤±2% F.S. |
| Repeatability | ≤1.0% |
| Zero Drift | ≤±2% F.S./h |
| Span Drift | ≤±2% F.S./3 h |
| Warm-Up Time | 30 minutes |
| Response Time | ≤60 s |
| Sampling Flow Rate | 0.5–2.0 L/min |
| Display Units | ppm and mg/m³ |
| Data Storage | 256 records |
| Communication Interfaces | RS485 and USB |
| Alarm | Audible and visual alarm |
| Sampling System | Built-in pump |
| Feature | ET-3015E | Conventional CO Monitoring Instruments |
| Measurement Principle | NDIR | Depends on sensor technology |
| Measured Parameters | CO, temperature, humidity | Often CO concentration only |
| Sampling Method | Built-in pump sampling | Depends on model |
| Display | Large color touchscreen | Varies by model |
| Concentration Units | ppm and mg/m³ | Depends on model |
| Automatic Zero Calibration | Supported | Model-dependent |
| Data Storage | 256 records | Model-dependent |
| Communication | RS485 and USB | Model-dependent |
| Alarm | Audible and visual | Model-dependent |
| Remote Sampling | Supports extended sampling lines | Depends on sampling configuration |