| Measurement mode | Vertical (vertical - upper or lower furnace body): Zero torque,zero friction, 180 ° linear measurement Automatic opening and closing of furnace body (equipped with high-end low-noise lifting system) More suitable for testing powder and paste samples (lower furnace design) Optional2-furnace system with free switching, expanding temperature range or improving efficiency |
| temperature range | -180-500℃ RT-1000/1400/1600/1650/1750℃ ≥ 2000 ℃ (specially customized) |
| Temperature control rate | 0-50/100 K/min(depending on the alloy SIC、 Precious metals, silicon molybdenum rods, graphite, etc 0-300K/min (light wave furnace) |
| refrigeration system | According to different furnace bodies, air cooling, water cooling, liquid nitrogen, etc |
| Temperature accuracy | 1K |
| Temperature Accuracy | 0.1K |
| Constant temperature technology for measurement systems | standard configuration |
| Sample holder | Quartz (customized by Heraeus) Aluminum oxide (customized by Morgan) Graphite (according to user requirements) |
| Number of samples | Single pole, double pole |
| sample length | 0<L≤50 mm |
| Sample diameter | 7 mm/12mm/14mm/20mm/Customized |
| measuring range | 0-5000 μm (±2500μm) |
| resolution | 0.03 nm/digit |
| Δ L/L0 repeatability | 0.001% absolute value |
| Accuracy of Δ L/L0 | 0.002% absolute value |
| measurement error | ± 1% (relative value) |
| Contact force | 0.01 N to2.5 N(auto loading) |
| vacuum degree | 10E-5 mbar |
| atmosphere | Oxidation, reduction, inert atmosphere, etc |
| gas control | Standard configuration: Manual gas box (2 channels) Optionalconfiguration: MFC (Bronkhorst) |
| water-cooling system | Standard configuration cooling function |
| vacuum system | Standard mechanical pump and accessories Optional: (According to user requirements) |
| Safety function | Softening point testing and protection Water flow switch technology Thermocouple damage protection technology |
| software | Includesworkstation and operationanalysis software; Sintering rate function RCS; Density calculation function; C-DTA calculation, etc; Compatible with Windows version. |