| model | BSC-1000B2 | BSC-1360B2 |
| Display control points | LCD display screen, wind speed display, pressure display | |
| suitable number of people | single person | two people |
| cleanliness level | Grade 100 (ULPA filter with a filtration efficiency of 99.9995% for 0.3um particles) | |
| Average inflow velocity (m/s) | 0.55 | 0.55 |
| Average flow velocity of descending airflow (m/s) | 0.33 | 0.33 |
| Control accuracy of descending airflow (m/s) | ±0.015 | |
| Noise level db | ≤67 | |
| Vibration half-peak value μm | ≤5 | |
| Average lighting Lux | ≥650 | |
| Ambient temperature ℃ | 10-30 | |
| Relative temperature (℃) | ≤70%(25℃) | |
| Atmospheric pressure Pa | 80-105 | |
| Personnel protection | When tested using the potassium iodide (KI) method, the protection factor of the front window operating port should not be less than 1×105 | |
| Protection of tested products | Colony count ≤ 5 CFU per instance | |
| Cross-infection prevention | The bacterial count of the impact sampler is ≤2 CFU per sampling | |
| Work area dimensions (length x width x height) mm | 1000X500X640 | 1300X500X640 |
| Overall dimensions (length X width X height) mm | 1120X895X2340 | 1420X895X2340 |
| Supply air filter specifications (length X width X height) mm | 1026X606X70 | 1326X606X70 |
| Exhaust air filter specification (length X width X height) mm | 482X515X90 | 782X515X90 |
| Fluorescent lamp specification W | 30 | 40 |
| Number of fluorescent lamps (pcs) | ||
| UV lamp specification W | 18 | 30 |
| Number of UV lamps (unit) | 1 | |
| Voltage V | AC 220V±10%/50Hz±2% | |
| Power (kW) | 2.1 | 2.2 |