  • Model Number3012H-D
    Brand NameQDLAOYING
    Payment TermsTT,PAYPAL,Alibaba.com Pay
    3012H-D Type High-Flow Low-Concentration Smoke Gas Analyzer
    Introduction:
    This instrument employs the Pitot tube parallel constant-speed sampling method to collect particulate matter from fixed pollution sources such as exhaust emissions. It uses the gravimetric filtration method to measure soot mass and applies constant potential electrolysis, UV differential, or non-dispersive infrared techniques for qualitative and quantitative analysis of flue gas composition. It is designed for measuring particulate matter emission concentrations, converted concentrations, and total emissions from various boilers, flues, industrial furnaces, and other fixed pollution sources, as well as for assessing the efficiency of dust removal and desulfurization equipment. The product is widely used in environmental protection, testing agencies, industrial enterprises (power plants, steel mills, cement plants, sugar mills, paper mills, smelters, ceramic factories, boilers, and furnaces, as well as industries such as aluminum, magnesium, zinc, titanium, silicon, pharmaceuticals, including chemical fertilizers, chemical plants, rubber, material factories, etc.), public health, labor safety, safety supervision, military, scientific research, education, and other fields.
    Features
    1. Capable of measuring ultra-low concentrations of particulate matter in flue gas from fixed pollution sources
    2 Pitot tube parallel isokinetic sampling method for collecting particulate matter from fixed pollution source exhaust;
    3. Real-time monitoring of working conditions such as dynamic pressure, static pressure, flue gas temperature, flow velocity, and other parameters;
    4. Precise electronic flow meter control with real-time temperature and pressure monitoring ensures flow accuracy;
    The system is equipped with sensors based on various principles such as constant potential electrolysis, UV differential method, and non-dispersive infrared method, allowing for autonomous sensor selection as needed to achieve simultaneous measurement of multiple gases;
    6. High brightness display with strong environmental adaptability; simple and user-friendly interface, intuitive data presentation.
    7. It features a self-diagnostic function for faults, enabling the detection of instrument malfunctions and providing alerts, which facilitates user maintenance and operation;
    The device can be optionally equipped with a built-in lithium battery pack or powered via an AC power adapter
    9 high-efficiency sampling pumps, corrosion-resistant, maintenance-free for continuous operation, suitable for various working conditions;
    10 It can be equipped with various sampling tubes as needed to achieve functions such as dust sampling, flue gas analysis, flue gas sampling, and direct reading of dust;
    Technical parameters:
    Sampling flow rate: (0~120) L/min
    Maximum sampling volume: 999,999.9 L
    Flue gas dynamic pressure: (0~2000) Pa
    Constant-speed tracking response time: no more than 20s
    Smoke static pressure: (-30~+30) kPa
    Sampling pump load capacity: ≥60 L/min (at a resistance of 20kPa)
    Pressure before the flow meter: (-60~0) kPa
    External dimensions: (Length × Width × Height) 290mm × 180mm × 275mm
    Temperature before the flow meter: (-55~125)°C
    Instrument noise: <80dB(A)
    Atmospheric pressure: (50~130) kPa
    Host weight: approximately 7.5 kg
    Flue gas temperature: (0~500)°C (extendable)
    Power consumption: <200W
    Dry and wet bulb temperatures (optional): (0~100)℃
    Flue gas flow rate: (1~45) m/s
    Power supply: Built-in lithium battery (22.2V/6Ah) (optional) or powered by an AC adapter (with built-in charging management)

     
     
  • Introduction:
    This instrument employs the Pitot tube parallel constant-speed sampling method to collect particulate matter from fixed pollution sources such as exhaust emissions. It uses the gravimetric filtration method to measure soot mass and applies constant potential electrolysis, UV differential, or non-dispersive infrared techniques for qualitative and quantitative analysis of flue gas composition. It is designed for measuring particulate matter emission concentrations, converted concentrations, and total emissions from various boilers, flues, industrial furnaces, and other fixed pollution sources, as well as for assessing the efficiency of dust removal and desulfurization equipment. The product is widely used in environmental protection, testing agencies, industrial enterprises (power plants, steel mills, cement plants, sugar mills, paper mills, smelters, ceramic factories, boilers, and furnaces, as well as industries such as aluminum, magnesium, zinc, titanium, silicon, pharmaceuticals, including chemical fertilizers, chemical plants, rubber, material factories, etc.), public health, labor safety, safety supervision, military, scientific research, education, and other fields.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com