Position: Home > > Environmental monitoring and analysis > Air sampling equipment > Particulate sampler > 2030D Smart Low-Flow TSPPM10PM2.5 Sampler

2030D Smart Low-Flow TSPPM10PM2.5 Sampler

Product ID:HQCYklw036

Price: Please inquire PDF Format
Supply Ability:50 SETS/MON
Port:SHENZHEN
  • Product Introduction
  • Other information
  • Model Number2030D
    Brand NameQDLAOYING
    Payment TermsTT,PAYPAL,Alibaba.com Pay
    2030D Smart Low-Flow TSP/PM10/PM2.5 Sampler 
    Introduction:
    This instrument is a multifunctional device capable of sampling TSP, PM10, PM2.5, microbial aerosols, as well as indoor inhalable and fine particulate matter, including benzo(a)pyrene, in the ambient atmosphere. It can be used for aerosol and particulate matter monitoring by departments such as environmental protection, health, labor, safety supervision, military, disease control centers, scientific research, and education.  
    Implementation standard:
    GB/T 16293-2010 Test Method for Airborne Microorganisms in Clean Rooms (Areas) of Pharmaceutical Industry
    GB/T 18203-2000 Hygienic Standard for Hemolytic Streptococcus in Indoor Air
    GB/T 18204.3-2013 Methods for hygienic examination of public places - Part 3: Airborne microorganisms
    GB/T 18883-2022 Indoor Air Quality Standard
    GB/T 38517-2020 General Principles for Sampling and Analysis of Particulate Biological Aerosols
    GB/T 39193-2020 Ambient Air - Determination of Particulate Matter Concentration - Gravimetric Method
    HJ 93-2013 Technical Requirements and Testing Methods for Ambient Air Particulate Matter (PM10 and PM2.5) Samplers
    HJ 194-2017 Technical Specifications for Manual Monitoring of Ambient Air Quality
    HJ 618-2011 Determination of PM10 and PM2.5 in Ambient Air - Gravimetric Method
    HJ 656-2013 Technical Specifications for Manual Monitoring Method (Gravimetric Method) of Ambient Air Particulate Matter (PM2.5)
    HJ 657-2013 Air and Waste Gas - Determination of Metallic Elements such as Lead in Particulate Matter - Inductively Coupled Plasma Mass Spectrometry Method
    JJF 1826-2020 Calibration Specification for Airborne Microbial Sampler
    Features:
    1. Wide flow range design, one machine for multiple purposes:
    It can be used for microbial aerosol sampling with Andersen's second, sixth, and eighth-stage sampling heads and impact absorption bottles
    It can be connected to a particulate matter sampling head for ambient air TSP/PM10/PM2.5 particulate matter sampling
    It can be applied to indoor air quality monitoring sampling, specifically for sampling inhalable particulate matter and fine particulate matter, including benzo[a]pyrene
    2. Equipped with a precision core pump, it boasts strong load-bearing capacity, making it ideal for high-load sampling
    3. Equipped with a high-capacity lithium battery, it is portable and allows for quick setup of sampling tasks
    4. The electronic flowmeter automatically and precisely controls the flow rate, with automatic control of sampling flow, ensuring stable flow
    5. Equipped with a large-capacity internal storage, the sampling data can be stored, viewed, exported, and printed
    6. Measure atmospheric pressure, ambient temperature, and other parameters in real-time.
    7. Equipped with a fault self-checking function, it can detect the functions of the instrument and indicate faults, facilitating user maintenance and usage;
    Technical parameters:
    Sampling flow rate: (5~50) L/min, operating point flow rates: 10.0 L/min, 16.67 L/min, 28.30 L/min
    Sampling time: set arbitrarily within 99 hours and 59 minutes
    Pre-measurement temperature: (-55~125)℃
    Pre-metering pressure: (-55~0) kPa
    Atmospheric pressure: (50~130) kPa
    Load capacity: At 28.3L/min, it can overcome a resistance of 22kPa
    Noise: ≤62dB(A)
    Dimensions: (Length × Width × Height) 334mm × 186mm × 254mm
    Weight: Without lithium battery: approximately 4.5kg. With lithium battery: approximately 4.7kg
    Power consumption: <50W
    Operating power supply: AC (220±22) V, 50Hz; built-in lithium battery (22.2V/3.5Ah, optional) or DC 24V

  • Introduction:
    This instrument is a multifunctional device capable of sampling TSP, PM10, PM2.5, microbial aerosols, as well as indoor inhalable and fine particulate matter, including benzo(a)pyrene, in the ambient atmosphere. It can be used for aerosol and particulate matter monitoring by departments such as environmental protection, health, labor, safety supervision, military, disease control centers, scientific research, and education.  
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com