HK-7000 gas alarm controller

Product ID:HQWRCO2CO023

Price: Please inquire PDF Format
Supply Ability:30 SETS/MONTH
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberHK-7000
    Brand NameHKY
    Payment TermsT/T, paypal
    Introduction:
    For users with more than 10 monitoring points, the HK-7000 gas alarm controller cabinet of our company's cabinet structure can be used. Its function is the same as that of the wall-hanging type, but the HK-7000 unit (11~70) can be installed centrally. The specific number of ways and cabinet structure can be customized according to user needs.
    Features:
    1. Meet the requirements of national standard GB16808-2008;
    2. Using the combination of master control card and channel card, a master card can carry up to 99 channel cards;
    3. The channel cards are designed independently, freely combined, and do not interfere with each other. They can be used with various 2-wire and 3-wire gas detectors.
    4. The channel card adopts 4 LED digital display and light bar trend display in two ways, intuitive and clear;
    5. Each channel card supports (4 ~ 20) mA output, and can be set independently of the two primary and secondary alarm output;
    6. The main control card has a black box function of history records, and the recorded contents include operations such as fault, alarm, power-on, power-off, and reset;
    7. The main control card supports RS-485 (Modbus) upload function, and contains a total alarm, the total fault passive contact output: silence, reset control contact input, easy access PLC/DCS and other industrial control systems;
    8. Has a password lock function to prevent misuse.
    Technical Parameters:
    Control loop: 1~99
    Operating environment: Temperature (0~55)°C, Humidity ≤93%RH (non-condensing)
    Input signal: (4~20)mA standard current signal, silence, reset control contact input
    Output signal: RS-485 bus and (4~20)mA, passive normally open contacts: AC220V/3A, DC24V/3A
    Measurement error: ±1%FS
    Alarm record: time, concentration, alarm hold, first alarm indication, total number of alarms
    Alarm function: Independent LED indication and different audible and visual signals at different levels indicate different alarm levels, and can latch alarm time and alarm status.
    Sound alarm: >80dB

  • Introduction:
    For users with more than 10 monitoring points, the HK-7000 gas alarm controller cabinet of our company's cabinet structure can be used. Its function is the same as that of the wall-hanging type, but the HK-7000 unit (11~70) can be installed centrally. The specific number of ways and cabinet structure can be customized according to user needs.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
18620722006
instruments@instrumentstrade.com