EFP110 closed mouth flash point tester Introduction
EFP110 automatic flash point tester uses a new high-speed digital signal processor, which has high reliability and accuracy. One host can control multiple test furnaces to test multiple samples at the same time, saving test time. Detection, opening, ignition, alarm, Cool, print, the whole experiment process is completed automatically;
Gas-electric dual-use ignition mode, namely platinum alloy heating wire and gas ignition mode;
Automatic detection of atmospheric pressure and automatic correction of test results, can be used in high altitude areas;
Newly developed high-power high-frequency switching power supply heating technology, high heating efficiency, adaptive PID control algorithm, automatic adjustment of heating curve;
Automatic detection of temperature over-value and alarm;
Thermal tail printer makes printing more beautiful and faster, with offline printing function;
Up to 500 historical records to be time-stamped;
Centennial clock with temperature compensation, accurate travel time, automatically record the date and event of measurement, can run for more than 10 years in the state of power failure;
Use 320 * 240 large screen graphic LCD display, Chinese character display interface, rich content;
Full touch screen keys for intuitive and convenient operation;
High degree of automation, can use electronic ignition and gas ignition two types, automatic ignition, automatic fire detection and fire detection, automatic detection and recording of flash point and storage, automatic printing report and other functions. Operation: Simple and convenient operation, safe and reliable.
The EFP110 closed mouth flash point tester is suitable for measuring the flash points of fuel oils, lubricating oils, solid suspensions, liquids that tend to form surface films and other liquids under experimental conditions. Features
1) Dual control system;
2) Double ignition mode;
3) Self-inspection and self-diagnosis;
4) Multiple interfaces;
5) The test process is intuitive;
6) Separate independent design;
7) True color display;
8) Gold-plated components;
9) Ignition test flash;
10) Overheating protection;
11) Test results are intuitive;
12) Multiple test procedures. Technical Parameters
1) Executive standard: GB / T261ISO2719ASTMD93
2) Flash point measurement range: room temperature to 370 ° C
3) Working voltage: AC220V, 50Hz
4) Overall power: 750W
5) Heating power: 630W
6) Working temperature: 0 ℃ ~ 40 ℃
7) Storage temperature: -20 ~ 50 ℃
8) Safety temperature: EFP + 20 ℃
9) Pre-safe temperature: EFP + 10 ℃
10) Absolute safe temperature: 400 ℃
11) Ignition mode: electronic ignition or gas ignition
12) Display: color dot matrix LCD, 6.4 ″
13) Display resolution: 640 × 480
14) Dimensions: 478 × 319 × 322
15) Weight: 25KG
The EFP110 closed-end flash point tester uses a new high-speed digital signal processor, which is reliable and accurate. The EFP110 closed-end flash point tester is suitable for measuring fuel oil, lubricants, solid suspensions, liquids that tend to form surface films and Flash point of other liquids.