Position: Home > > Environmental monitoring and analysis > Water organic pollutants > TOC analyzer > CU-500S Online Total Organic Carbon Analyzer for Ultra-pure Water Preparation System in the Semiconductor Industry

CU-500S Online Total Organic Carbon Analyzer for Ultra-pure Water Preparation System in the Semiconductor Industry

Product ID:HSYJtoc022

Price: Please inquire PDF Format
Supply Ability:20 SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberCU-500S
    Brand NameYIMA
    Payment TermsT/T, paypal
    CU-500S Online Total Organic Carbon Analyzer

    Introduction
    CU-500S total organic carbon analyzer use Static UV oxidation-Conductivity Detection method to detect the concentration of total organic in de-ionized water, range from 0.1ppb-1000ppb. It has a high sensitivity and accuracy, as well as good stability.
    Application
    Can be used for online monitoring the pharmaceutical industry water system, ultra- pure water preparation system in the semiconductor industry, and the deionized water preparation process in power plants.
    Fundamentals
    Using UV light oxidation of carbon in water samples, after changes in the conductivity before and after oxidation of the sample to obtain the corresponding TOC values. During the test, water samples were isolated alone in a closed space, thus preventing contact with the surrounding environment pollution caused. The TOC value is through changes in the conductivity of the water samples before and after oxidation by UV light to obtain the corresponding operations. The conductivity of the water sample is directly related to the content of organic carbon in water samples, and organic carbon content and can be completely oxidized to CO2 and measured.
    Product Features:
    1. High accuracy
    2. Lower test limit
    3. Sense temperature and conductivity while detect organic carbon concerntraction
    4. 4-20mA Interface can be connected
    5. 3.5 Inch touch screen
    6. Data convenient query
    7. Five years of data storage
    8. UV lamp life Countdown
    9.Embedded dot matrix printer
    10. Low maintenance costs
    11. No reagents and carrier gas
    12. Compact size, light weight
    13. Wall hanging design, easy online installation

    Technical Parameters:
    Detect range: 0.1-1000.0μg/L 0.1ppb
    Accuracy error: ≤±5%
    Repeatability error: ≤±3%
    Span drift: ≤±2%/D
    Zero drift: ≤±2%/D
    Conductivity detection range: 0.055-6.000μS/cm
    Showing Data: TOC, Conductivity, temperature
    Response time: Less than 15mins
    Injection method: Water network pressure
    Reagent: No need
    Communication: 4-20mA(External)
    Display and operator: 3.5 inches screen
    Data storage: 5-years data (date measured 500 times)
    Options: Calibration Pump
    Power: 220V AC≤±10% 50Hz
    Output: 60W
    Ambient temperature: 0-40 degree Celsius
    Dimensions (W*D*H mm): 305*126*168
    Weight: 6kg

     

  • Introduction
    CU-500S total organic carbon analyzer use Static UV oxidation-Conductivity Detection method to detect the concentration of total organic in de-ionized water, range from 0.1ppb-1000ppb. It has a high sensitivity and accuracy, as well as good stability.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
18620722006
instruments@instrumentstrade.com