The light source | Semiconductor laser, life is more than ten years > 30000 h |
Sampling flow | 100 LPM/min (3.53 ft3) import sampling pump plus or minus 5% |
Display mode | 122 x 90 mm high resolution LCD display (LCD) in both Chinese and English interface, real-time display, |
Power supply | The AC power, AC, 220 v + 5%, 50 HZ |
Working time | 10 hours |
Counting mode | The accumulative total value, difference, density maximum sampling concentration: 35000 / l |
The test way | Single, repetitive, continuous, calculation, remote, |
Unit conversion | Units can be converted to m3, ft3 95% UCL computing, can directly show the particle concentration (stars/stere, star/cubic feet) |
Sampling period | 1 ~ 9999 (S) delay counting: 0 ~ 99 (S) from a net time: 10 or less (min) |
Temperature and humidity: |
The choose and buy: (1) temperature 0 ℃ to 50 ℃ ~ + 1. (2) the humidity: 0 ~ 100% RH + / - 5% |
UCL set |
Sampling points (A) : 2 ~ 9 set. Each point of sampling number: (L) 2 ~ 9 times set measuring location: 0-999 |
UCL statements |
ISO14644 2001-1, GB50073-2001, the new version of GMP, FS - 209 |
Communication interface | RS232, 9600 baud rate |
Overlapping error | 5%, 2000000 grain/cubic feet |
Print function | The built-in printer |
Weight | 12.8kg |
Overall dimensions | 280×400×230 |